datasheetbank_Logo
数据手册搜索引擎和 Datasheet免费下载 PDF

ADAS1000(RevA) 查看數據表(PDF) - Analog Devices

零件编号
产品描述 (功能)
比赛名单
ADAS1000 Datasheet PDF : 80 Pages
1 2 3 4 5 6 7 8 9 10 Next Last
Data Sheet
ADAS1000/ADAS1000-1/ADAS1000-2
NOISE PERFORMANCE
Table 3. Typical Input Referred Noise over 0.5 Second Window (µV p-p)1
Mode
Data Rate2
GAIN 0 (×1.4)
±1 VCM
Analog Lead Mode3
High Performance Mode
2 kHz (0.5 Hz to 40 Hz)
8
2 kHz (0.05 Hz to 150 Hz) 14
GAIN 1 (×2.1)
±0.67 VCM
6
11
GAIN 2 (×2.8)
±0.5 VCM
5
9
GAIN 3 (×4.2)
±0.3 VCM
4
7.5
1 Typical values measured at 25°C, not subject to production test.
2 Data gathered using the 2 kHz packet/frame rate is measured over 0.5 seconds. The ADAS1000 internal programmable low-pass filter is configured for either 40 Hz or
150 Hz bandwidth. The data is gathered and post processed using a digital filter of either 0.05 Hz or 0.5 Hz to provide data over noted frequency bands.
3 Analog lead mode as shown in Figure 58 and Figure 57.
Table 4. Typical Input Referred Noise (μV p-p)1
Mode
Analog Lead Mode3
High Performance Mode
Low Power Mode
Electrode Mode4
High Performance Mode
Low Power Mode
Digital Lead Mode5, 6
High Performance Mode
Low Power Mode
Data Rate2
2 kHz (0.5 Hz to 40 Hz)
2 kHz (0.05 Hz to 150 Hz)
16 kHz
128 kHz
2 kHz (0.5 Hz to 40 Hz)
2 kHz (0.05 Hz to 150 Hz)
16 kHz
128 kHz
2 kHz (0.5 Hz to 40 Hz)
2 kHz (0.05 Hz to 150 Hz)
16 kHz
128 kHz
2 kHz (0.5 Hz to 40 Hz)
2 kHz (0.05 Hz to 150 Hz)
16 kHz
128 kHz
2 kHz (0.5 Hz to 40 Hz)
2 kHz (0.05 Hz to 150 Hz)
16 kHz
2 kHz (0.5 Hz to 40 Hz)
2 kHz (0.05 Hz to 150 Hz)
16 kHz
GAIN 0 (×1.4) GAIN 1 (×2.1) GAIN 2 (×2.8) GAIN 3 (×4.2)
±1 VCM
±0.67 VCM ±0.5 VCM
±0.3 VCM
12
8.5
6
5
20
14.5
10
8.5
95
65
50
39
180
130
105
80
13
9.5
7.5
5.5
22
15.5
12
9
110
75
59
45
215
145
116
85
13
9.5
8
5.5
21
15
11
9
100
70
57
41
190
139
110
85
14
9.5
7.5
5.5
22
15.5
12
9.5
110
75
60
45
218
145
120
88
16
11
9
6.5
25
19
15
10
130
90
70
50
18
12.5
10
7
30
21
16
11
145
100
80
58
1 Typical values measured at 25°C, not subject to production test.
2 Data gathered using the 2 kHz packet/frame rate is measured over 20 seconds. The ADAS1000 internal programmable low-pass filter is configured for either 40 Hz or
150 Hz bandwidth. The data is gathered and post processed using a digital filter of either 0.05 Hz or 0.5 Hz to provide data over noted frequency bands.
3 Analog lead mode as shown in Figure 58 and Figure 57.
4 Single-ended input electrode mode as shown in Figure 59 and Figure 57.
5 Digital lead mode as shown in Figure 59 and Figure 57.
6 Digital lead mode is available in 2 kHz and 16 kHz data rates.
Rev. A | Page 9 of 80